Kwetsbaarheden worden geladen…
Kwetsbaarheden worden geladen…
CVE-2026-7251
The affected product is vulnerable to due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.
Dit record: live koppeling — laatst opgehaald: 25 juli 2026 om 01:21.
Leverancier
Eppendorf
Product
BioFlo 320 Bioreactor
Gepubliceerd
25 juli 2026
Laatst gewijzigd
25 juli 2026
The affected product is vulnerable to due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Attack vector
NETWORK
Privileges required
NONE
User interaction
NONE
Vertrouwelijkheid
Geen
Integriteit
Geen
Beschikbaarheid
Geen
Beperkte impact op beschikbaarheid; risico ligt vooral bij vertrouwelijkheid of integriteit van procesdata.
Geen sterke energiesector-specifieke signalen herkend; algemene OT/ICS-relevantie.
Eppendorf has released a software update that permanently removes VNC access from the controller. Users should download and apply this update from: https://www.eppendorf.com/software-downloads.